Basic technical information of pressure transmitter
Nominal pressure range | 0÷10kPa to 0÷60MPa |
---|
Overpressure | 200% (400%) nominal range |
---|
Error | max. 1% (0.5%) |
---|
Zero temperature error | typ. 0,1 max. 0,3 %/10°C |
---|
Span temperature error | typ. 0,1 max. 0,3 %/10°C |
---|
Compensated temp. range | 20 ÷ +85°C |
---|
Operating temp. range | -15 ÷ +85°C (non condensing) |
---|
Storage temperature | -25 ÷ 100°C |
---|
Supply voltage | 12 ÷ 36V dc |
---|
Supply voltage – output H | < 4mA |
---|
Output | 4 ÷ 20mA two-wire 0 ÷ 20mA three-wire 0 ÷ 10 V three-wire |
---|
Operated position | arbitrary |
---|
Protection | min. IP 54 |
---|
Voltage strength case - electronics | min. 1000 V dc |
---|
Weight | cca 200g |
---|
EMC – according to ČSN EN 61326-1
At the customer´s wish it is possible to secure a metrological verification of the transmitters at an accredited Calibration Service Centre.
Instructions for use
- Before connection of the transducer into the pressure circuit, it is necessary to verify that the pressure being measured corresponds to the nominal range of this transducer. Even a transient loading over the maximum allowable overpressure may cause a destruction of the measuring diaphragm!
- Measured media is in contact with stainless steel, Viton sealing and ceramic membrane. If you use aggressive media, it is necessary to verify the transducer material resistance.
- Transmitter with open membrane must be sealed by the “O” ring on six-edge surface to avoid mechanical strain in sensor fitting.
- In the case of turning-off with connector nut you must avoid to turning of connector base ! In opposite case input wires can be ruptured !
- The device requires no maintenance during operation. Potential reparation performs manufacturer. Device will be electronic waste after end of its usage. The user is obliged to guarantee liquidation of device in accordance with a valid prescription at the liquidation time. It is recommended deliver device to authorised person.